;UoM_Project ;T2 measurement using Periodic Refocusing of J evolution by Coherance Transfer (PROJECT) using perfect spin echo ; ;Developed By NMR Group ;Department of Chemistry ;University of Manchester ;United Kingdom ;2019 ; ;J.A. Aguilar, M. Nilsson, G. Bodenhausen, G. Morris, Chem. Commun.,2012,48,811-813. ;$CLASS=HighRes ;$DIM=1D ;$TYPE= ;$SUBTYPE= ;$COMMENT= #include #include "p2=p1*2" "d11=30m" "acqt0=0" baseopt_echo 1 ze 2 d11 pl1:f1 d1 p1 ph1 3 d20 p2 ph2 d20 p1 ph3 d20 p2 ph2 d20 lo to 3 times c go=2 ph31 d11 mc #0 to 2 F1QF(ivc) exit ph1=0 2 0 2 1 3 1 3 2 0 2 0 3 1 3 1 ph2=1 1 1 1 0 0 0 0 3 3 3 3 2 2 2 2 ph3=1 1 3 3 0 0 2 2 3 3 1 1 2 2 0 0 ph31=0 2 0 2 1 3 1 3 2 0 2 0 3 1 3 1 ;pl1 : f1 channel - power level for pulse (default) ;p1 : f1 channel - 90 degree high power pulse ;p2 : f1 channel - 180 degree high power pulse ;d1 : relaxation delay; 1-5 * T1 ;d11: delay for disk I/O [30 msec] ;d20: fixed echo time to allow elimination of J-mod. effects ; d20 should be << 1/J ,but > (50 * P2) [1-2 msec] ;vc : variable loop counter, taken from vc-list ;NS: 2 * n, total number of scans: NS * TD0 ;DS: 2 ;define VCLIST ;this pulse program produces a ser-file (PARMOD = 2D) ;vc : vc should contain even numbers to provide ; for cancellation of 180 degree pulse errors